Sector 74, Mohali, Punjab
MOQ : 1 Piece
Business Type : Manufacturer, Exporter, Supplier, Trader, Importer
Product Details
Type | HPLC System |
Accuracy - Pump | +/- 1% or +/- 2µl/min, whichever is greater// +/- 1% or +/- 1µl/min, whichever is greater (Micro) |
Flow Rate Settings - Pump | 0 – 0.10 mL/min. 0.001 inc.//0.10 – 0.90 mL/min. 0.010 inc.//1.0 mL/min. + 0.10 inc. |
Precision - Pump | 0.25% from 0.1 mL/min to 10 mL/min. at 20oC//0.25% from 0.05 mL/min to 4 mL/min. at 20oC (Micro) |
Pmax – Pump | Pmax |
System Connections - Pump | 1/16” Capillaries |
Dimensions – Pump | 410 x 180 x 270 mm. |
Weight – Pump | 7.3 Kgs. |
Power Input | 85 to 264V AC, 47 to 63 Hz, 60 Watts. |
Display | 320 mm. |
Control | Standalone mode using front panel |
Payment Terms | NEFT, IMPS, DD, D/P, T/T, Other |
Brand | VSI |
Additional Information
Product Code | VSI-3201 |
Port | MOHALI |
Delivery Time | Ready Stock or within 1-2 weeks |
Preferred Buyer From
Location | Worldwide |
VSI Electronics Private Limited is an established manufacturer, supplier and exporter of HPLC System from India. HPLC System offered by the company are high on quality, performance and are made of the finest quality raw materials, making them durable and long lasting.
HPLC System (High Pressure Liquid Chromatography System), VSI-3201 is an isocratic system with high pressure solvent delivery pump, Multi-wavelength UV Detector, Sample inject valve Rheodyne 7725i. LC Column (USA) with work-station software. Wavelength range 190-700nm, Wavelength accuracy +/-1nm, spectral bandwidth 8nm, flow cell volume 8 micro litre, optical length – 10nm.
SALIENT FEATURES
Note: Appearance & Specifications are subject to change due to continuous development.
Technical Specifications Of HPLC System
Wavelength | 190-700 nm. |
Spectral Bandwidth | 6 nm |
Wavelength Accuracy | <= +/- 1 nm. |
Wavelength Reproducibility | <= +/- 0.5 nm. |
Flow Pool Volume | 10 µl |
Optical Path | 10 mm |
Dynamic Noise | </= 2.5 x 10-5 AU (Methanol, 1 ml/min, 254nm) |
Static Noise | </= + 1 <10 -5 AU (Empty pool, Response Time 1s) |
Dynamic Drifting | </= + 2.5 x 10 -4 AU (Methanol, 1 ml/min, 254nm) |
Static Drifting | </= + 1 x 10 -4 AU (Empty pool, Response Time 1s) |
Qualitative Repeatability | RSD </= 0.1% |
Quantitative Repeatability | RSD </= 0.5% |
Dimensions – Detector | 160H X 260W X 350D (mm.) |
Weight – Detector | 6 Kg. |
Minimum Detection Concentration | </= 1 x Square root 10 -8 g/mL (Naphthalene/Methanol Solution) |
Standard Accessories | Rheodyne Injector 7725 I (USA)// Micro Syringe// Column C-18// Software CD// Filters// Instruction Manual |